AGRICULTURE
MDR Technology (Multi Data Rate) in the 2.4 GHz range
The MDR (Multi Data Rate) technology makes it possible to transmit high data rates and large data sets in the shortest possible time. The spread spectrum transmission and a multiple transfer of the records ensure transmission reliability and a fast and trouble-free measurement process, as one expects from m&h.
AFS Technologie (Automatic Frequency Select)
With AFS (Automatic Frequency Select) technology, the frequency ranges of this radio band are continually scanned and partial frequencies are automatically selected which are free of interfering signals. This not only ensures the system's own transmission with be fast and trouble-free, but also prevents interference form WLAN systems or other radio sources.
Unidirectional Repeatability (deflection from one direction) | max. 1 μm (2 Sigma) with 50 mm stylus and 254 mm/min probing feedrate |
Recommended Probing Feedrate | Max. 2000 mm/min |
Trigger Force with 50 mm Stylus | PP41.00 - XY = 0.3 - 1.4 N; Z = 2.5 - 12.5 N MY21.00 - XY = 1 N; Z = 6 N |
Maximum Stylus Overtravel | PP41.00 - XY ±12,5°; Z - 6 mm MY21.00 - XY ±14°; Z - 4,5 mm |
Sensing Directions | ±X, ±Y, -Z |
Transmission Frequency | 433 MHz / 2.4 GHz |
Transmission/Reception Range | Up to 18 m |
Power supply | 1x 9 V battery, block, 6LR61, Lithium: 1200 mAh, Alkaline: 550 mAh |
Material | Stainless steel, POM |
Weight without Shank | approx. 920 g |
Temperature Range | Operation: 10°C - 50°C, Storage: 5°C - 70°C |
Sealing | IP68: EN60529 |